Vibration Isolation Case Studies
DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.
Series
Equipment Type
Manufacturer
All Case Studies

JEOL Tabletop SEM
Installation of DVIA-T56 Active Vibration Isolation Platform for JEOL JCM-7000 Tabletop SEM

ZIESS-GeminiSEM 500 DVIA-MB1000 (P165 / 200522R2-2) Installation report — 2021.02.01
DVIA-MB1000 worksheet P165 (serial 200522R2-2) for ZIESS-GeminiSEM 500 (measurement dated 2021.02.01; reporting table 2021.01.29; operator Choi, HyoungMun, Daeil System). Overview, transmissibility summary table, and axis screenshots follow the DOCX; MB1000 internal-program captures align with Z, X, and Y headings per §2a embed mapping. Place of installation, END USER, and place of measurement were blank in the source and are published as N/A. The decorative raster on the centred MB1000 title line is not republished.

HITACHI Regulus 8230 DVIA-MB1000 Program (P173 / 191219R3-4) Installation Report — 2021.01.29
MB1000 program internal handout (Word title MB1000_P173 191219R3-4): overview, transmissibility summary at 2 / 5 / 10 Hz for Z/X/Y with a short qualitative note, and MB1000 internal-program Z/X/Y performance plots. The centered graphic inserted immediately under the Word title line is omitted as cover decoration; the source file has no separate Reference section or VC comparison figures.

BOE ZEISS Crossbeam 550 DVIA-MB1000 (MB1000 P181, 200720R5-1) Internal Program — 2021.01.22
DVIA-MB1000 worksheet (MB1000 P181, platform 200720R5-1) for BOE with target equipment ZEISS Crossbeam 550. Setup and reporting 2021-01-22; operator Choi, HyoungMun (DAEIL SYSTEMS research engineer). Transmissibility summary at 2 / 5 / 10 Hz and Z / X / Y MB1000 internal-program plots under the axis captions, plus the setup photo from the report title block.

ZEISS Crossbeam 550 DVIA-MB1000 P188 (200720R5-3) Internal Program Performance — 2021.01.22
DVIA-MB1000 internal program handout (platform P188, reference 200720R5-3) for ZEISS Crossbeam 550. Reporting date 2021-01-22; setup and measurement 2021-01-21 by Choi, HyoungMun (DAEIL SYSTEMS). Publishes transmissibility summary values from the form and Z / X / Y internal-program screenshots mapped per DOCX §2a. Decorative title-band raster (image1) is omitted from the public gallery; END USER and installation / measurement site fields are N/A in the source.

Zeiss Crossbeam 550 DVIA-MB1000 (MB1000_P182 / 200720R5-2) Internal Program Performance — 2021.01.22
DVIA-MB1000 (MB1000_P182) with reference 200720R5-2, target equipment ZEISS Crossbeam 550. Transmissibility summary at 2 / 5 / 10 Hz and MB1000 internal-program screens follow the DOCX §2a head-to-embed map (Z / X / Y). The title-band picture under the report heading is omitted from the public gallery.