Vibration Isolation Case Studies
DAEIL SYSTEMS engineers active vibration isolation systems for the world's most demanding nanoscale imaging and metrology environments.
Series
Equipment Type
Manufacturer
All Case Studies

Tekscend Photomask CLB KLA Reticle Inspection System KLA597XR DVIA-P4000 Installation Report
DVIA-P4000 (serial 241202R4) tuning and vibration measurement at Tekscend Photomask CLB for a KLA KLA597XR Reticle Inspection System. Engineer Chaewon Lee (DAEIL SYSTEMS); measurement and report December 29, 2024; B&K LAN-XI DAQ and PCB 393B05. Equipment not installed at measurement time. Published vibration specification capture and vertical, left-to-right, and front-to-back velocity spectral density plots.

QRT DVIA-T67 DEMO (241213R2) Inspection Report — MSTech MST8500 Probe Station
December 2024 DEMO shipment tuning and vibration report for DVIA‑T67 (serial 241213R2) at Korea Institute of Nano Technology 3F QRT lab — MSTech MST8500 probe station. Measurements with equipment Turned off; summary table cites Table vs DVIA‑T67; Brüel & Kjær Pulse Labshop acquisition; VC / Autospectrum / Transmissibility by axis plus equipment photo. Reference VC criteria as markdown only.

Self-made - DVIA-M1000 (241010R2) Installation Report
DVIA-M1000 installation report (serial 241010R2) — Self-made - -. Field package. UI-program transmissibility plots and equipment photo; Reference as markdown VC table.

Seoul National University DVIA-T56 (130123R3) Installation Report
DVIA-T56 (serial 130123R3) tuning and vibration measurement at Samsung Leaders Building Room 1104 — level foot rework, equipment-not-installed condition; VC curves, autospectrum, and transmissibility by axis.
Tier-1 Semiconductor Cheonan C5 ZEISS Crossbeam 550 FIB-SEM DVIA-ML1000 (240710R2) Installation Report
DVIA-ML1000 installation report (serial 240710R2) for ZEISS FIB-SEM Crossbeam 550 at Tier-1 Semiconductor Cheonan C5. Overview describes on-site tuning and vibration measurement with the tool in IDLE after placement; measurement setup and equipment vibration limit VC-E appear in the specification capture. Measurement Summary includes a vertical floor line FAIL with platform PASS; horizontal axes PASS. Published plots: vibration specification capture, VC / Autospectrum / Transmissibility by axis. Reference uses the standard Generic Vibration Criteria markdown table only (no separate reference graphic).

Ciqtek High Speed SEM HEM6000 DVIA-M1000 (241101R6) Installation Report
DVIA-M1000 installation report (serial 241101R6) — Ciqtek High Speed SEM HEM6000. Field package. UI-program transmissibility plots and equipment photo; Reference as markdown VC table.