Amazon Web Services Elionix EBL DVIA-M3000 (210915R2) Installation Report — 2022.04.22
DVIA-M3000 (serial 210915R2) installation report dated 22 April 2022 for Elionix EBL at Amazon Web Services (form "Prepared for: AWS"). VC and transmissibility curves for vertical, left-to-right, and front-to-back axes on floor and active isolator, vibration results summary table, and Generic Vibration Criteria as a markdown Reference table. Decorative header bitmap and the standalone Reference criteria figure from the handout were omitted in favor of the site table.

Carl Zeiss Magok LG Science Park ZEISS Versa 620 (X-ray) DVIA-MB3000 (220310R3) Installation Report — 2022.04.22
DVIA-MB3000 (serial 220310R3) worksheet from 22 April 2022 at Carl Zeiss Magok LG Science Park with Versa 620 (X-ray): summary VC table versus floor and active-platform states for Z-, X-, and Y-axis VC curves plus transmissibility plots. Engineers Youngha Lee and Seounghoon Cheon (DAEIL SYSTEMS); report written date 22.04.25. Reference VC criteria are published only as an on-site markdown table (no reference figure asset).

Hanyang University DVIA-MB1000 (141211R5) Installation Report — 2022-04-21
Compact installation report prepared for Hanyang University with DVIA-MB1000 (serial 141211R5) on SEM equipment at the Sungdong, Seoul campus: overview, vibration summary comparing floor versus active VC-class bands across Z / X / Y, VC curves plus transmissibility for each axis at floor versus active measurement points (report header decoration bitmap omitted from publication), and the standard VC reference markdown table.

Korea Institute of Science and Technology (KIST) Microscope DVIA-UD350 (220127R1-1) Installation Report
DVIA-UD350 (SN 220127R1-1) at KIST Building L7 Room 7223, Seongbuk-gu, Seoul, supporting a microscope. Engineers Youngha Lee and Seounghoon Cheon (DAEIL SYSTEMS); tuning and report dated April 21, 2022. Measurement summary table plus Z/X/Y VC curves (Floor and Active), transmissibility plots, and VC criteria as markdown (header decoration image and Word Reference raster omitted).

Korea Institute of Science and Technology (KIST) Microscope DVIA-UD350 (220127R1-2) Installation Report
DVIA-UD350 (SN 220127R1-2) at KIST Building L7 Room 7223, Seongbuk-gu, Seoul, supporting a microscope. Engineers Youngha Lee and Seounghoon Cheon (DAEIL SYSTEMS); tuning and report dated April 21, 2022. Measurement summary table plus Z/X/Y VC curves (Floor and Active), transmissibility plots, and VC criteria as markdown (header decoration image and Word Reference raster omitted).
Tier-1 Semiconductor ZEISS Versa 620 DVIA-MB3000 (220118R8) Installation Report — 2022.04.18
DVIA-MB3000 (serial 220118R8) installation worksheet with tuning on April 18, 2022: VC curves and transmissibility for floor vs active platform (Z, X, Y), summary VC band table, and Reference as a markdown Generic Vibration Criteria table (no separate reference figure). Engineers Youngha Lee and Seounghoon Cheon (DAEIL SYSTEMS); report written 22.04.25.