1264 case studies
Thermo Fisher Scientific Apreo 2S HiVac DVIA-M1000 Installation Report

GRGT / Shanghai — Apreo 2S HiVac on DVIA-M1000 (230905R2). M1000 internal program measurement (full view and Z/X/Y plots).

TUSUR, Tomsk Emcraft Veritas 3100 DVIA-MB1000 Installation Report — 2023.10.27

DVIA-MB1000 (SN 220825R6-3) under Emcraft Veritas 3100 for TUSUR, Tomsk, Russia — engineer itinerary, onsite tuning window, Z/X/Y transmissibility plots (no standalone equipment photograph in source).

Guangdong Jiashang Solaris GMH DVIA-M1000 (230809R2) Installation Report

DVIA-M1000 abbreviated worksheet (serial 230809R2). Setup dated 23.11.02 and reporting dated 23.11.03; measurement date 2023.11.02 (Thursday) in Dongguan with operator Jongwon Park from DAEIL SYSTEMS. END USER Guangdong Jiashang New Energy Tech Co., Ltd.; target equipment Solaris GMH and installation site Dongguan as in the form. The Overview block states the standard DVIA-M1000 sentence only. No vibration specification, equipment status, tuning request, measurement summary table, or Reference section appears in the source. The internal-program section title in the source reads “M1000 internal program performance measurement result”; published figures are three plots labeled Z axis, X axis, and Y axis. Decorative header graphics from the Word file are omitted from public copy.

Huawei Solaris GMH DVIA-M1000 Installation Report (230509R6-2)
Installation Report
DVIA-M Series
11-01-2023

Huawei Solaris GMH DVIA-M1000 Installation Report (230509R6-2)

Huawei-end Solaris GMH site (Shanghai): DVIA-M1000 serial 230509R6-2 setup — internal program performance screenshots for Z, X, and Y axes.

Sungkyunkwan University JEOL JBX-8100FX DVIA-P2200 (220509R3) Inspection Report

DVIA-P2200 (serial 220509R3) inspection at Sungkyunkwan University Research Building for JEOL JBX-8100FX electron beam lithography. Engineers Youngha Lee and Jongwon Park; tuning and report October 31, 2023. Feedback/feedforward re-tuning completed with good performance; Caution AIR LED from low regulator pressure (no performance impact). Summary VC grades (floor vs active) and Z/X/Y VC curves from the inspection handout.

Dow Chemical Sigma 300 DVIA-M1000 (230328R3) Installation Report

DVIA-M1000 abbreviated worksheet for serial 230328R3 (source filename M1000_230328R3_23.10.31.docx). Setup and reporting dated 23.10.31; measurement date 2023.10.31 (Tuesday) in Shanghai with operator Sangjae Lee (DAEIL SYSTEMS). END USER Dow Chemical; target equipment Sigma 300 and installation site Shanghai as in the form. The Overview block states the standard DVIA-M1000 sentence only. No vibration specification, equipment status, tuning request, measurement summary table, or Reference section appears in the source. The internal-program section title reads “M1000 internal program performance measurement result”; published figures are three plots labeled Z axis, X axis, and Y axis. The form header line shows (M1000_ 230328R3 Set up). Decorative header graphics from the Word file are omitted from public copy.