69 Ministry of Defense Thermo Fisher Sceintific Prisma Installation Report
Contents
Overview
The DVIA-ULF700 active vibration isolation platform is appropriately installed and functioning as intended.
Vibration Isolation System Information
Model: DVIA-ULF700
Serial Number: 251023R8
Engineer
Chaewon Lee from DAEIL SYSTEMS
Tuning Date
December 11, 2025
End User
69 Ministry of Defense
Number of Tuning Trial
1st tuning
Location
Ho chi Minh, Vietnam
Equipment
Manufacturer: Thermo Fisher Scientific
Equipment: SEM
Model: Prisma
Equipment Condition
The equipment is turned on/IDLE
Measurement Summary
| Measurent Point | Status | AXIS | Measurement Data | |
|---|---|---|---|---|
| Floor | DVIA-ULF700 | |||
| 1. Floor 2. DVIA-ULF700 | Prisma installed/turned off | Vertical | VC-D @ 3.15 Hz | VC-G @ 3.15 Hz |
| Left to Right | VC-E @ 3.15 Hz | VC-G @ 3.15 Hz | ||
| Front to Back | VC-E @ 2 Hz | VC-G @ 2 Hz | ||
Measurement Data Obtained via UI Program
Vertical VC Curves
Vertical Narrow Band
Vertical Transmissibility
Left to Right VC Curves
Left to Right Autospectrum
Left to Right Transmissibility
Front to Back VC Curves
Front to Back Autospectrum
Front to Back Transmissibility
Equipment Picture
Reference
Generic Vibration Criteria
Notes:
1. VC-A/B is measured in 1/3 octave bands from 8-80 Hz, VC-C through VC-G from 1-80 Hz.
2. Detail size refers to line widths in microelectronics manufacturing or particle sizes in medical research.